Abstract:
A new method for the determination of trace amounts of cysteine has been reported. The method is base on measurement of the substace produced by the reaction of cysteine of the substace produced by the reaction of cysteine with 12-molybdosilicate,which can be strongle absorbed on DME and produced a anodic wave at -9.6v(Vs.SCE) in NH 3-NH 4Cl buffer solution (pH=9.6). The derivative peak height is proportional to the concentration of cysteine in the range of 1×10 -7  ̄5×10 -5 mol/L.The detection lim...